Tenmars TM-288 HEAT STRESS WBGT Meter

TENMARS

Regular price $159.95


The chance of heatstroke increases dramatically in the moist heat environment.
This product, measuring in WBGT, provides real-time alarm immediately so as to
prevent heatstroke in such the environment.

 

Functions:

  • In compliance with JIS B 7922 Level 2
  • Three applications: arm band, tripod, and hanging.
  • High-sensitive capacitive humidity sensor and bandgap temperature sensor.
  • Measure the WBGT, black-globe temp. (TG), Air temp. (TA), humidity (%RH).
  • Low voltage indicator
  • °C or °F selectable.
  • With the plastic black-globe of the diameter Ø40mm.
  • Three-stage WBGT alarm selectable:     mute (buzzer not displayed), cyclic setting.
  • Outdoor or indoor (default is OUT) measurement mode selectable.
  • Sampling interval: per 20 seconds
  • Operating wind speed: 0.3 m/s or above.
  • Operation temperature and humidity: 0°C to +50°C,<80%RH (No condensation)
  • Storage temperature and humidity: 0°C to +60°C,<70%RH (No condensation)
 

    Wet Bulb Globe Temperature (WBGT)

    Unit

    Measuring range

    Alarm Default

    Alarm User
    Flashing
    Caution >21~25°C

    mute

    As exceeding the setting value,beep for 15 seconds

    Extreme caution >25~28°C

    Beep for 4 seconds

    Danger >28°C~31°C

    Beep for 15 seconds

    Extreme Danger >31°C

    Beep for 3 second and pause for 2 seconds, repeat until 15 seconds


    Wet Bulb Globe Temperature (WBGT)

    Unit

    Measuring range

    Resolution

    tolerance range

    15~40°C

    0~50.0 0.1 ±2.0
    32.0~122.0 0.1 ±3.6

    Indoor & outdoor without sunshine ”IN” :WBGT = (0.7×WET)+(0.3×TG)

    Outdoor in sunshine ”OUT” :WBGT=(0.7×WET)+(0.2×TG)+(0.1×TA)


    Air temperature (TA)

    Unit

    Measuring range

    Resolution

    Tolerance

    @15~40°C

    0~50.0 0.1 ±0.6
    32.0~122.0 0.1 ±1.6

    Black-globe temperature (TG)

    Unit

    Measuring range

    Resolution

    Tolerance

    @15~40°C

    0~80.0 0.1 ±0.6
    32.0~176.0 0.1 ±1.1

    Humidity (%RH)

    Unit

    Measuring range

    Resolution

    Tolerance

    @25°C

    %RH 30.0~90.0 0.1 ±5.0%RH
    %RH <30.0 , >90.0 0.1 ±7.0%RH